| Mechanical
Dimensions |
|
4.75”W
x 3”H x 1.8”D |
| Mechanical
Weight |
|
Internal
pump models: < 16 oz. Diffusion models: <
15 oz |
| Operating
Temperature Range |
|
-20°C
to +50°C (-4°F to +122°F) |
| Relative
Humidity Range |
|
0 to
98% (non condensing) |
| Power Sources |
|
Rechargeable
NiCad battery pack or 2 AA sized alkaline batteries
|
| Battery
Life |
|
Internal
pump models: 8 hours. Diffusion models: 10 hours
|
| Alarm Outputs |
|
Audible:
> 90 dB at 1’ Visual: Red LED, flashing
display characters |
| Approvals |
|
UL
classified and CSA certified Class I, Division 1,
Groups A, B, C, D Class II option–UL classified
|
| Warranty |
|
Lifetime.
Excludes sensors, batteries and pump |
| Repeatability |
|
+ 2%
|
| |
|
|
| Gas Ranges
& Alarm Set Points |
|
Sensor
|
Instrument
Range |
Immediate
Alarm |
Accuracy |
| |
|
Combustible
(LEL) |
0-100%
LEL |
10%
LEL |
2%
LEL or
5% FS |
| |
|
Oxygen
(O2) |
0-25%
by volume |
19.5%(low)23.5%(high) |
0.5%
by volume |
| |
|
Carbon
Monoxide (CO) |
0-800
PPM |
35
PPM(25 PPM in CA) |
+
10% of reading |
| |
|
Hydrogen
Sulfide (H2S) |
0-500
PPM |
10
PPM |
+
10% of reading |
| |
|
Sulfur
Dioxide (SO2) |
0-20
PPM |
2
PPM |
+
10% of reading |
| |
|
Hydrogen
Cyanide (HCN) |
0-800
PPM |
5
PPM |
+
10% of reading |
| |
|
Phosphine
(PH3) |
0-20
PPM |
0.3
PPM |
+
10% of reading |
| |
|
Ammonia
(NH3) |
0-500
PPM |
25
PPM |
+
10% of reading |
| |
|
Chlorine
(Cl2) |
0-100
PPM |
0.5
PPM |
+
10% of reading |
| |
|
Chlorine
Dioxide (ClO2) |
0-10
PPM |
0.1
PPM |
+
10% of reading |
| |
|
Combo
Tox (CO/H2S) |
0-800
PPM |
35/15
PPM |
+